A1 40 DTR.APC.APR.ALW.03
(ENG)
LCD2DP setting the decimal point position on LCD2. Options:
XXXXX·
XXXX·X
XXX·XX
XX·XXX
X·XXXX
In a situation where the value provided to display on the display LCD2 can not be
displayed properly due to the position of the decimal point, this is indicated by displaying
the four flashing dots • • • •. In this case, you must enter to the local menu setting and
move the decimal point respectively to the right.
FACTOR
back to factory settings (removing of current and pressure calibrations).
MID_WP blocking records / change the parameters associated with transmitter metrology
ON switch on blocking
OFF switch of blocking
10.2.6. Local Menu, error reports.
During executing in Local Menu some functions, LCD2 announcement can be displayed on the screen.
The error displaying evidences about no realization of command of Local Menu.
The shortened description of errors announcements is showed below.
ERR_L07 [in_write_protected_mode]. Error will ensue out when we try to change setting in Local
Menu, but transmitter is protected before recording. To make the change of setting with
Local Menu using, transmitter has to have the included service of Local Menu as well as
switched off protection before record. These parameters modification is possible by using
KAP -03 communicator, “RAPORT” program or software using library EDDL.
•
default setting:
Local Menu service switched on
protection before record switched off
ERR_L09 [applied_process_too_high]. Error will ensue out when given parameter (pressure)
exceed admissible value. Zeroing or the range setting verifying is necessary.
ERR_L10 [applied_process_too_low]. Error will ensue out when given parameter (pressure)
will too low. Zeroing or the range setting verifying is necessary.
ERR_L14 [span_too_small]. Error will ensue out when in result of setting range executing change
the width of the range will be smaller than admissible.
ERR_L16 [acces_restricted]. Error will ensue out when the service of Local Menu is switched off,
and the user tries to call out the Menu Local service. You should switch on the service of
Local Menu with the KAP-03 communicator, “RAPORT” program, or software using
library EDDL.
Warning!, ERR_L16 announcement can be displayed as well by zeroing attempt of the
absolute transmitter .
WNG_L14 [WARNING!, new Lower Range Value Pushed !] Error will ensue out when the end of
range set ( the URV) change will cause the change of the range set beginning (LRV).
10.2.7. Remote configuration
Remote configuration is possible with KAP-XX communicator or PC software. Measuring circuit should be in
accordance with the fig. 2.