EasyManuals Logo

AquaLab 4TE Series User Manual

AquaLab 4TE Series
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #107 background imageLoading...
Page #107 background image
16 APPENDIX B AquaLab
16 Appendix B
Temperature Correction
of Decagon’s Verification Standards
Table 5: Water Activity of Selected Salt Solutions
Temp. H
2
O
0.50 2.33 6.00 8.57 13.40
(
C)
mol/kg mol/kg mol/kg mol/kg mol/kg
KCL NaCL NaCL LiCl LiCl
15.0 1.000 0.984 0.923 0.761 0.492 0.238
20.0 1.000 0.984 0.922 0.760 0.496 0.245
25.0 1.000 0.984 0.920 0.760 0.500 0.250
30.0 1.000 0.984 0.920 0.760 0.504 0.255
35.0 1.000 0.984 0.920 0.760 0.508 0.261
40.0 1.000 0.984 0.921 0.760 0.512 0.266
50.0 1.000 0.984 0.894 0.740 0.517 0.275
Note: AquaLab will measure these verification standards to ±0.003
a
w
.
102

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TE Series and is the answer not in the manual?

AquaLab 4TE Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 to 50 °C
Measurement Time5 minutes
Measurement PrincipleChilled Mirror Dewpoint
Resolution0.0001 aw
Temperature Accuracy±0.1 °C
CommunicationUSB, RS-232
DisplayLCD
Measurement Range0.03 to 1.00 aw
Operating Temperature4-50 °C (39-122 °F)
Data Storage8000 readings
Power Supply50/60 Hz
Storage Temperature-20 to 50 °C

Related product manuals