AquaLab 17 APPENDIX C
17 Appendix C
AquaLab Verification Standards
Application Note
Using AquaLab is easier then ever. Pre-packaged standard salt solu-
tions are immediately available for performance verification, saving
you time and money. Validation and documentation for GMP and
GLP has also become easier. Operate your instrument with cer-
tainty and insure the quality of your food product by using low cost
precision salt solutions.
No need to purchase and store reagent grade salts.
No additional laboratory equipment necessary.
Avoid solution handling and mixing errors.
Save technician time.
The AquaLab should be verified against a known salt standard daily.
For high use or batch processing, the instrument should be checked
regularly against a known salt standard of similar water activity.
Checking the water activity of a standard solution will alert the op-
erator to the possibility of contamination of the unit or shifts in the
linear offset from other causes.
Now, you can verify AquaLab performance with confidence. Per-
formance Verification Standards come in five water activity levels;
0.984, 0.920, 0.760, 0.500, and 0.250 a
w
. The standards are pro-
duced under a strict quality assurance regime by an independent
third party that verifies the standards and they are shelf stable for
one year. Order your calibration salt standard of similar water ac-
tivity today.
Uncertainties Using Saturated Salt Solutions
The water activity values listed in our operator’s manual for sat-
urated salts were reprinted from Greenspan (1977). His method for
determining water activity was to combine all of the available data
from tests by other researchers. He did not set up any experiments of
103

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TE Series and is the answer not in the manual?

AquaLab 4TE Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 to 50 °C
Measurement Time5 minutes
Measurement PrincipleChilled Mirror Dewpoint
Resolution0.0001 aw
Temperature Accuracy±0.1 °C
CommunicationUSB, RS-232
DisplayLCD
Measurement Range0.03 to 1.00 aw
Operating Temperature4-50 °C (39-122 °F)
Data Storage8000 readings
Power Supply50/60 Hz
Storage Temperature-20 to 50 °C