4 GETTING STARTED AquaLab
4 Getting Started
4.1 Components of your AquaLab
Your AquaLab should have been shipped with the following items:
AquaLab water activity meter
Calibration certificate
Power cord
USB interface cable
50 disposable sample cups
Operator’s Manual
Quick start guide
Cleaning kit
Two vials each of the following verification solutions:
1.00 a
w
USP Purified Water
0.984 a
w
0.50 mol/kg KCL
0.920 a
w
2.33 mol/kg NaCl
0.760 a
w
6.00 mol/kg NaCl
0.500 a
w
8.57 mol/kg LiCl
0.250 a
w
13.41 mol/kg LiCl
4.2 Choosing a Location
To ensure that your AquaLab operates correctly and consistently,
place it on a level surface. This reduces the chance that sample ma-
terial will spill and contaminate the sample chamber. Also select a
location where the temperature remains fairly stable to avoid tem-
perature changes that can affect accuracy. This location should be
well away from air conditioner and heater vents, open windows, etc.
Place the AquaLab in a location where cleanliness can be maintained
to prevent contamination of the sample chamber.
14

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TE Series and is the answer not in the manual?

AquaLab 4TE Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 to 50 °C
Measurement Time5 minutes
Measurement PrincipleChilled Mirror Dewpoint
Resolution0.0001 aw
Temperature Accuracy±0.1 °C
CommunicationUSB, RS-232
DisplayLCD
Measurement Range0.03 to 1.00 aw
Operating Temperature4-50 °C (39-122 °F)
Data Storage8000 readings
Power Supply50/60 Hz
Storage Temperature-20 to 50 °C