5 MENUS AquaLab
Temperature
The default temperature is 25
C. Press the Enter button to change
the temperature setting. The AquaLab Series 4TE models may be
set between 15 and 50
C by 0.1
C intervals. Using the up and down
arrows, set the AquaLab to your desired temperature and press the
save button.
Temp Eq
The Temperature Equilibration option allows you to set the level
of temperature equilibration desired before the water activity mea-
surement begins. The range is 0.5 to 4.0
C. A setting of 4.0
C
begins the measurement immediately (assuming the sample is not
> 4.0
C above or below the block temperature). A setting of 0.5
C will cause the instrument to wait until the sample temperature is
within < 0.5
C of the block temperature before starting the water
activity measurement.
18

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TE Series and is the answer not in the manual?

AquaLab 4TE Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 to 50 °C
Measurement Time5 minutes
Measurement PrincipleChilled Mirror Dewpoint
Resolution0.0001 aw
Temperature Accuracy±0.1 °C
CommunicationUSB, RS-232
DisplayLCD
Measurement Range0.03 to 1.00 aw
Operating Temperature4-50 °C (39-122 °F)
Data Storage8000 readings
Power Supply50/60 Hz
Storage Temperature-20 to 50 °C

Related product manuals