AquaLab 12 TROUBLESHOOTING
be made. (Refer to Chapter 8, Samples Not at Room Temper-
ature.)
3. Some products absorb or desorb moisture very slowly, causing
measurements to take longer than usual, and nothing can be
done to speed up the process. Refer to Chapter 8 for further
explanation.
4. Your sample may contain volatiles. Volatiles are known to
cause unstable readings, because they condense on the surface
of the chilled mirror and alter readings. Please refer to the
volatiles section in Chapter 8 for hints on reducing difficulties
with measuring samples with propylene glycol. If you have fur-
ther questions regarding the measurement of volatiles contact
Decagon.
5. A fan blade in the block chamber may be broken or bent. If
even salt standards take a long time to read, and the sample
chamber is clean, you may have a broken chamber fan blade.
This is especially likely if you have just cleaned the chamber.
If you suspect this may have happened, contact Decagon for
details on replacement.
3. PROBLEM:
Water activity readings on verification standards are too high/low
and a linear offset adjustment cannot be made any higher/lower.
SOLUTIONS:
1. The thermopile in your chamber, which measures sample tem-
perature, may have become contaminated. Refer to Chapter 6
for directions on cleaning.
2. The chamber mirror may be dirty. Refer to Chapter 6 for di-
rections on cleaning.
4. PROBLEM:
Message on screen displays a warning that the sample is too hot:
67

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AquaLab 4TE Series and is the answer not in the manual?

AquaLab 4TE Series Specifications

General IconGeneral
Accuracy±0.003 aw
Temperature Control15 to 50 °C
Measurement Time5 minutes
Measurement PrincipleChilled Mirror Dewpoint
Resolution0.0001 aw
Temperature Accuracy±0.1 °C
CommunicationUSB, RS-232
DisplayLCD
Measurement Range0.03 to 1.00 aw
Operating Temperature4-50 °C (39-122 °F)
Data Storage8000 readings
Power Supply50/60 Hz
Storage Temperature-20 to 50 °C

Related product manuals