3-17
W/Var/PF
a) W/Var/PF Meter Calibration Connections,
no Current or Potential Transformers
b) W/Var/PF Meter Calibration Connections,
with Current and Potential Transformers
W/Var/PF
W/Var/PF/PA
Inst.
Pwr.
+
-
Output
W/Var/PF/PA
Inst.
Pwr.
+
-
Output
c) W/Var/PF/PA Transducer Calibration Connections,
no Current or Potential Transformers
d) W/Var/PF/PA Transducer Calibration Connection
with Current and Potential Transformers
Figure 6
Single-Phase Watt, Var, or Power Factor Meters and
Watt, Var, Power Factor, or Phase Transducers, Calibration Connections
RB RB
1040C
out
V
out
Current
Connections
Voltage
Connections
Current
Connections
Voltage
Connections
R
B
B
R
1040C
out
V
out
1040C
out
V
out
RBRB
Current
Connections
Voltage
Connections
B R
1040C
out
V
out
R
B
Current
Connections
Voltage
Connections