2-43
Voltage Settings
RANGE FULL SCALE OUTPUT MODIFY VALUE OUTPUT VALUE
1 100 mV 2600 15340
2 1 V 2600 15340
3 10 V 2500 14750
4 100 V 2600 15200
5 1000 V 2600 15200
Upon completion of the current output amplifier offset test, use the MODIFY UP or MODIFY
DOWN keys to indicate pass or fail respectively. The operator may repeat the Debug Utilities or
select other Diagnostic Tests.
2.9 Error Messages
The PMC displays numerous error messages to warn the operator of improper entries and setups,
out-of-bound conditions, and internal faults. Along with the message display the errors are
reported by code number on the Hand Held Unit display and over the GPIB. Table 2.10 lists the
various error messages as seen on the display with a description, probable cause, appropriate
action to be taken, and the corresponding code number. The PMC may always be returned to
normal operation by pressing the RESET key.
A failure during the Function Test or one of the Diagnostic Tests is reported with FAIL appearing
in the last four display locations. If a test failure persists consult a repair technician.

Table of Contents

Other manuals for Arbiter Systems 1040C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arbiter Systems 1040C and is the answer not in the manual?

Arbiter Systems 1040C Specifications

General IconGeneral
BrandArbiter Systems
Model1040C
CategoryTest Equipment
LanguageEnglish