1-5
Table 1.1 - Current Specifications (continued)
____________________________________________________________________________
TYPE DESCRIPTION
____________________________________________________________________________
II. Output, AC Current
Range 4 0.1 A to 1.05 A rms
Range 5 1 A to 7.5 A rms
AC Current Amplitude +(0.2% of reading +0.05% of full scale)
Accuracy +(0.2% of reading +0.1% of full scale) on Range 5
Compliance 6 V rms
Overload Setting 6.5 V rms
Distortion 0.45% Max
Frequency 50 Hz to 75 Hz and 333.3 Hz to 500 Hz
Frequency Accuracy <0.01% of reading
Current Output Stability <(0.03% of reading +0.015% of full scale);
AC and DC averaged 1 minute or longer
Resolution and Setability <0.1% of reading
Settling Time <8 seconds Max