2-49
Table 2.10 - Error/Probable Cause/Appropriate Action (continued)
___________________________________________________________________________________________________________
CODE DISPLAYED
# MESSAGES DESCRIPTION PROBABLE CAUSE APPROPRIATE ACTION
___________________________________________________________________________________________________________
9 CAL. STORAGE
ERROR
Whenever new calibration
constants are stored a check
sum is created. Upon
power-up the constants are
read and the check sum is
verified. This error
indicates the check sum
verification failed.
Possible faulty component on
processor card.
Turn PMC off and back on again.
If problem persists consult repair
technician.
10 CAL. LIMIT
EXCEEDED
The calibration adjustment
limit has been exceeded.
During calibration the output has
been adjusted too far in one
direction.
Check measurement equipment to
verify calibration value. If the
problem persists an internal fault
is indicated - consult repair
technician.
ENTRY ERROR An out-of-bounds value or
incorrect units were input.
Incorrect key strokes. After a few moments the error
message will disappear and
normal operation may continue.
ERROR Generic If failure persists, consult repair
technician.