3-6
involve circuit connections which are similar,
if not identical to those employed by single
phase watt and power factor meters and
transducers. Therefore, representative circuit
connections and calibration connections for
these devices are shown in figures 5 and 6,
respectively. When connected per the
calibration setup diagram, calibration of these
instruments can be performed in accordance
with the meter manufacturer's
recommendations. For more detailed
instructions on operating the 1040C in the
volt-amps and volt-amps reactive modes, refer
to section 2.5.6 of the operation manual.