7.0 Technical Description
7.1 Basic Technical Specifications
Maximum deviation from fixed factors
±5% (unless otherwise noted)
–5 to +40°C (+23 to +104°F)
10% to 80%, non-condensing
Storage and transportation
–20 to +60°C (–4 to +140°F)
Classification / Spec Compliance
IEC 60601-1 (Amend 1 & 2), 60601-1-3,
60601-2-7; 21 CFR 1020.30 & 1020.31
MDD (93/42/EEC): Annex IX
IPX0; do not operate under wet conditions
Conductive connection to patient
No conductivity with the applied part
For use in environments where no flammable anesthetics and/or flammable cleaning agents are
present; non-acetone based disinfectant wipes or cloths.
Rechargeable lithium batteries
22.2V nominal; 25.2V maximum, 1.25A/hr
Low battery alert set point
Any cell (of the 6) <3.5V
Battery current at 2.5mA, 60kVp output
Maximum energy input in one hour
X-ray Controls and Generator
0.02–1.00 sec. (in 0.01 second increments)
1:60 (one 1 second exposure every 60 seconds)
Minimum permanent filtration
≥1.5mm Al (0.8mm glass, 0.5mm Al, 0.2mm plastic cap)
150W nominal at 60kV, 2.5mA
2.5mA (±5%) at 60kVp (±10%)
Leakage technique factors
Maximum air kerma at handgrips and
control panel