7.4 EMC Data
Independent laboratory testing for electromagnetic compatibility for conformity to the Sub-clause 5.10
requirements of ISO/IEC 17025 "General Criteria for the Competence of Testing and Calibration
Laboratories" was conducted by:
Nemko USA, Inc.
11696 Sorrento Valley Rd, Suite F
San Diego, CA 92121-1024
USA
The NOMAD Pro Dental X-ray System has been tested and found to comply with the limits of
electromagnetic compatibility standards for medical devices, which provide reasonable protection
against harmful interference in a typical medical/dental setting. The NOMAD Pro may generate and
radiate radio frequency energy that causes interference to other devices in the vicinity, if not used in
accordance with the instructions (though there is no guarantee that interference will not occur in a
particular instance). If interference occurs, the user is encouraged to try the following corrective
measures: reorient or relocate the receiving device; increase the separation between the equipment;
consult the device manufacturer or field service technician for help.
NOMAD Pro is intended for use in the electromagnetic environment as specified. The following tables
describe the tests performed and the status of the testing. The NOMAD Pro uses RF energy only for its
internal function. In addition, the housing includes conductive shielding; therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
The test results show that the NOMAD Pro is suitable for use in all establishments, including domestic
establishments and those directly connected to public low-voltage power supply network that supplies
buildings used for domestic purposes.
Test Specifications: Radio Frequency Emissions and Electromagnetic Immunity tests in accordance
with requirements of EN 60601-1-2:2001, A1:2006 (as follows).
Test Type
In Accordance with
Document
Document Title
Conducted and Radiated
Emissions
EN 55011:2007,
Class "B"
Industrial, Scientific and Medical (ISM) Radiofrequency Equipment – Radio
Disturbance Characteristics – Limits and Methods of Measurement
Electrostatic Discharge
Immunity
IEC 61000-4-2:1995,
A1:1998, A2:2000
Electromagnetic Compatibility, Basic Immunity Standard, Electrostatic
Discharge Immunity Test
Radio Frequency Immunity IEC 61000-4-3:2006
Electromagnetic Compatibility, Basic Immunity Standard, Radiated Radio
Frequency Electromagnetic Field, Immunity Test
Power Frequency Magnetic
Field Immunity
IEC 61000-4-8:1993,
A1:2000
Electromagnetic Compatibility, Testing and Measurement Techniques for
Power Frequency Magnetic Field, Immunity Test