NOTE: The tests documented in the table above are the only tests required for this product as it is a
battery operated device and the AC Adapter (CP-0062) is certified. IEC 61000-3-2, 3-3, 4-4, 4-5, 4-6,
4-11, and EN 55014-1:2006 are not applicable.
Emissions Test Summary
EN 55011:2007, Group 1, Class "B"
Conducted Emissions
0.15MHz – 30.00MHz PASS
EN 55011:2007, Group 1, Class "B"
Radiated Emissions
30.0MHz – 2500MHz PASS
EN 61000-3-2:2000, A2:2005
Power Line Harmonics
Up to the 40th Harmonic N/A, for a battery powered device
EN 61000-3-3:1995, A1:2001, A2:2005
Power Line Flicker
Less than or equal to 4% Maximum Relative Voltage
&KDQJH9DOXHRI'7IRUPRUHWKDQPV
N/A, for a battery powered device
EN 55014-1:2006
Discontinuous Disturbance ("Click")
0.15MHz – 30.00MHz N/A, for a battery powered device
Immunity Test Summary
Specification
Minimum Test Level Required per
EN 60601-1-2
Test Level Completed
IEC 61000-4-2:1995, A1:1998,
A2:2000 – Electrostatic Discharge
Immunity
Air discharge up to ±8kV; contact
discharge up to ±6kV
Air discharge up to ±8kV; contact
discharge up to ±6kV
PASS
IEC 61000-4-3:2006 – RF
Radiated Fields Immunity
Radiation field strength of 3V/m from
80 – 6000MHz (80% AM @ 1kHz)
Radiation field strength of 3V/m from
80 – 6000MHz (80% AM @ 1kHz)
PASS
Corrigendum 1:2006 – Electrical
Fast Transient Immunity
Power line pulses of ±2kV direct; I/O
line pulses of ±1kV
Power line pulses of ±2kV direct; I/O
line pulses of ±1kV
N/A, for a battery
powered device
IEC 61000-4-5:2005 – Lightning
Power line surges of ±2kV common,
Power line surges of ±2kV common,
IEC 61000-4-6:2003, A1:2004,
A2:2006 – RF Common Mode
Immunity
150kHz – 80MHz at 3 Vrms, 1kHz
80% amplitude modulated
150kHz – 80MHz at 3 Vrms, 1kHz
80% amplitude modulated
N/A, for a battery
powered device
IEC 61000-4-8:1993, A1:2000 –
Power Frequency Magnetic Field
Immunity
Helmholtz coil at 50Hz and 60Hz, to
3 amps (rms) per meter
Helmholtz coil at 50Hz and 60Hz, to
3 amps (rms) per meter
PASS
IEC 61000-4-11:2004 – Voltage
Dips and Short Interruptions
Voltage Dips of >95%, 30% and
60%; Interruptions of >95%
Voltage Dips of >95%, 30% and
60%; Interruptions of >95%
N/A, for a battery
powered device