Other
English
Recommended separation distances between
portable and mobile RF communications equipment and the EJ-MLA26 Series
The EJ-MLA26 Series is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the EJ-MLA26 Series can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the EJ-MLA26 Series as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.21.22.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
d 1.2 P=
d 1.2 P=
d 2.3 P=
EJ-MLA26_EN_Arthrex.book Page 39 Wednesday, August 8, 2012 11:58 AM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex EJ-MLA26UA and is the answer not in the manual?

Arthrex EJ-MLA26UA Specifications

General IconGeneral
BrandArthrex
ModelEJ-MLA26UA
CategoryMonitor
LanguageEnglish