EasyManuals Logo

Arthrex EJ-MLA26UA User Manual

Arthrex EJ-MLA26UA
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #5 background imageLoading...
Page #5 background image
5
English
Introduction
Do not use the monitor in the following
locations
In environments with a high oxygen content in the
atmosphere
In proximity to anesthetic gas (i.e. in a flammable
atmosphere)
In an atmosphere containing oxidized materials
Points of caution and points to note for
use
This monitor is designed to display images from a
camera etc. in a clinical environment.
It should not be used for diagnostic purposes.
When installing and using this monitor, it can generate
energy at radio frequencies. When not used and
installed according to the Operating Instructions, the
monitor can affect other devices. If the monitor cause
interference, which can be determined by turning the
other devices off and then on, the user is encouraged
to try to correct the interference by one or more of the
following measures:
Reorientate or relocate the monitor.
Increase distance between the monitor and other
devices.
Equipment connected to this monitor must conform to
IEC standards (i.e. IEC 60601-1 for Medical electrical
equipment, IEC 60950-1 for Information technology
equipment).
In addition, the combination and installation shall
conform to IEC 60601-1.
Any person who connects this monitor is responsible
for ensuring that the system conforms to
IEC 60601-1-1.
To conform to IEC 60601-1, equipment connected to
this monitor should be connected to the power supply
via an isolation transformer and have at least basic
insulation.
When connecting to other devices, there is a risk that
the leakage current may increase.
If there is anything that you do not understand, please
contact the dealer where the monitor was purchased.
Exercise adequate caution with regards to
static electricity or when using devices
nearby that emit electromagnetic waves
When using devices that emit electromagnetic waves
near the monitor, or when static electricity is present, it
is possible that the image on the monitor will
experience interference.
In such a case, please eliminate the effect by changing
the placement and orientation of the device etc.
EJ-MLA26_EN_Arthrex.book Page 5 Wednesday, August 8, 2012 11:58 AM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arthrex EJ-MLA26UA and is the answer not in the manual?

Arthrex EJ-MLA26UA Specifications

General IconGeneral
BrandArthrex
ModelEJ-MLA26UA
CategoryMonitor
LanguageEnglish