-5-
2. Basic Operations
CAUTION
Operate away from direct sunlight at a stable temperature with as little
fluctuation as possible.
Acclimate both the instrument and sample solution to room temperature to
ensure accurate results. Measurements may fluctuate with hot or cold samples.
Wait to press
F
after placing the sample on the prism. Alternatively, press
F
multiple times.
Operate on a flat work surface.
The displayed temperature is that of the prism and may not necessarily match
the temperature of the sample.
1. Zero-setting
This is the operation to set the calibration point (zero) for optical rotation and Brix.
m
Zero-set the instrument at the beginning of each day before use.
m
Regular zero-setting throughout the day is recommended to ensure accurate results.
Press I to turn it on.
m
Optical rotations and Brix can be
zero-set in any scale, however, when
the optical rotation scale is selected,
even when Brix zero-set is an error, no
error message will display.
Wipe the sample stage clean.
Fill the sample stage with tap water up
to the fill line.
N
Use room temperature tap water. The
instrument should also be acclimated
to room temperature prior to
operations.
Press G.
 flashes and stops.
N
To perform zero-setting on the optical
rotation scale and check whether Brix
zero-setting was done correctly, press
K and check that 000 is displayed.
Fill line
(Approx. 3mL)