EasyManua.ls Logo

ATAGO RePo-1 - Error Messages

ATAGO RePo-1
27 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
-14-
6. Error Messages
The battery power is low.
Not enough light could pass through
the sample solution to take the angle
of rotation measurement.
a
Dilute or filter the sample before
measuring.
An error has occurred whil
e
adjusting the setting.
Too much external light is interfering
with the measurement.
a
Shade the sample stage with
your
A hand and measure again.
*
This error message may also
appear if the sample temperature
is too low.
The sample measured above
the measurement range.
Measurement error.
(When User scale is selected)
a
Press J and K to
A view the detail of the error.
The sample measured below
the measurement range.
(Example)
The sample measured below 3.9˚C
(16.9˚C in the Purity or ISS scale).
G was pressed with
nothing or something other
than water on the sample
stage. (When Brix or User
scale is selected)
(Example)
The sample measured above
41.1˚C.















Related product manuals