FG_F6_PROFINET_EN_04 / 2020-06-17
Last results
Last results structure
At the end of each cycle, the last result is as an array of 40 words. This result includes the final state of
the instrument (relays position, alarm signal, indicators state…), but also of the test (units, values
measured for the pressure and the flow).
The last result is in the memory of the instrument. To obtain them, it is necessary to carry out a “Read
last results” request.
Image of the relays:
Bit 0 = 1: pass part.
Bit 1 = 1: fail part, maximum flow reject.
Bit 2 = 1: fail part, minimum flow reject.
Bit 3 = 1: alarm.
Bit 4 = 1: unused.
Bit 5 = 1: reserved.
Bit 6 = 1: unused.
Bit 7 = 1: unused.
Alarm code (refer to the alarm codes table).
Pressure unit code low part word (refer to units table).
Pressure unit code high part word (refer to units table).
Leak unit code low part word (refer to. Units table).
Leak unit code high part word (refer to. Units table).
Pressure piezo 2 low part word.
Pressure piezo 2 high part word.
Pressure piezo 2 unit code low part word (refer to units table).
Pressure piezo 2 unit code high part word (refer to units table).
Test check result low part word.
Test check result high part word.
Test check result unit code low part word (refer to units table).
Test check result unit code high part word (refer to units table).
Large Leak low part word.
Large Leak high part word.
Large Leak unit code low part word (refer to units table).
Large Leak unit code high part word (refer to units table).