4 Connections and Insulation Block Diagram 
© Copyright Reserved Autonics Co., Ltd.    33 
4.3  Wiring precautions 
  Mixing up the input terminals with output terminals and vice versa can lead to product 
damage. 
  Use only sensors supported by the product. 
  Make sure to connect rated SSRs or loads to the output terminals. 
  Make sure to connect the communication cable with correct communication terminals (A, B). 
  Make sure to observe correct polarity of power source terminals. (+ and -). 
  Use crimp terminals of size specified below. 
 
 
 
1 to 12  10mm 
Max. 1.7mm  Max. 3.7mm 
13 to 22  10mm 
Max. 2.1mm  Max. 4.2mm 
 
4.3.1  Sensor connection 
(1)  Compensation wire connection 
For thermocouple sensors, use compenstion wire of the same specification as input sensors. 
Using an extension wire of different specification and/or material will increase inaccuracy of 
temperature sensing. It is recommended to choose high performance compensation wire for 
more reliable sensing. 
(2)  Measurement error 
  Do not mix up the direction of the input sensor connector. 
  Carefully adjust both load and sensor positions. 
  Make sure the sensor is securely attached to the input connector. 
(3)  Wiring with AC power lines 
Do not put the sensor lines in close proximity of the AC power lines. 
 
 
  Make sure the sensor is completely inserted in the connector using a crimped terminal. 
  Fix the sensor to the connecter properly for accurate measurement. 
b