To avoid that the wet microtome is cooled
down and consequently the guides freeze
solid, the error code E-03 is shown on the
display. Now the microtome together with the
knife carrier must be dried.
The user is to make sure that the microtome is
in a perfectly dry condition. Then press the
memo button (fig. 2.12) to start the cooling
process. The error code E-03 goes off and the
cooling process starts.
The error code E-04 is shown on the left
display if the rear end position of the integrated
feed system is not reached within 2 min. after
turning on the instrument. The instrument can
be operated, however, the rear end position is
not defined. For this, press the memo button.
The error code
E-04 goes off.