EasyManua.ls Logo

Avantik QS11 - 5-2 Removing the microtome

Avantik QS11
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Avantik Cryostat QS11 / QS11UV
Avantik
36 Commerce St.
Springfield NJ 07081 / USA
387 779 - English
58
5-2 REMOVING THE MICROTOME
In the left rear part of the microtome there is
the so-called connector housing including
one electrical connection.
Bring the handwheel handle into its lower
position, i.e. the specimen clamping is also
in its lower position.
Unscrew the Allan screw (fig. 25.1) in the
middle of the front part of the microtome
base plate via an Allen screwdriver size
5 mm.
Caution:
It is recommended for all further work
to wear isolating gloves because of
the danger of freezer burns on the frozen
microtome.
Move the microtome to the left side until the
stop. This way, the clutch on the right side
of the microtome becomes free.
Move the microt
ome forwards and remove
it
from the cryo-chamber.
Thoroughly clean and wash the interior.
Caution:
Chloric cleaning agents must not be
used for cleaning purposes. In case of
a malfunction hazardous gases might
be generated together with the cooling brine
Tyfoxit of the cooling system.
Pull off the red stopper at the bottom of the
chamber. The cleaning liquid is collected in
the container (fig. 26.1) with a volume of
4,8 l at the front lower side
Note:
Please observe the filling level!!
The red stopper must be inserted after
cleaning as otherwise cold air leaks out
during cooling operation. Thus generating
frost built-up.
The installation and re-setting up of the
instrument is carried out in reverse order.
Fig. 25
Fig. 26
3
1
2
1

Table of Contents

Other manuals for Avantik QS11