For transportation outside closed
buildings, please observe the following
measures:
Turn off the instrument.
Unplug the unit.
Remove blade and/or knife from blade
and/or knife carrier.
Remove section waste trays, brush shelf,
blade/knife carrier and other accessories
as well as tools from the cryo chamber.
Note:
Clean and disinfect these accessories
according to the respective applicable
lab regulations and transport them in
dry condition.
Clean and disinfect the cryo chamber
according to the respective applicable lab
regulations.
Unscrew the handwheel handle for
transportation.
Note:
To make sure that these is no
condensation water inside the
chamber,
let the instrument approx.
48
h dry.
Loosen the fastening screws of the rollers
to move the instrument.
Arrange the dismounted accessories into
the intended transportation packing.
Note:
Before the further transportation,
tighten the fastening screws so that the
instrument does not move during
transportation.
Slightly tilt the cryostat and insert the foam
parts underneath it.
Place the case over the instrument.
Insert the wooden cover.
Attach the tightening strips around the
packing.
Caution
:
Any shipping of the instrument requires original
packaging materials! Damages caused by
shipping with non-conform packaging are not
covered by the manufacturer warranty! Any
damage repairs resulting in non-conforming
package are fully charged to the sending party.
We reserve the right depending on
seriousness of damage NOT to repair.
To order original packaging materials, please
contact Avantik Biogroup LLC or your local, by
Avantik Biogroup LLC authorized, dealer.
Caution:
The user must care for a clean and safe
condition of the instrument when returning
it to an appropriate service provider.
Note:
If the original packaging is no longer available,
please contact your local Avantik Biogroup
LLC representation.