5-7
Chapter 5. ADVANCED OPERATION
z
Parameter setup item list
No. Display Description
Factory
Setting
Setting Range
Referential
function setup
Remarks
1
0.r
G
*
1
Flowrate OK
judgment range
(2 % FS)
*
11
(0.5 to 100 % FS)
*
11
C-07
,
C-08
Unit: L / min(standard)
2
0.HyS
*
1
Flowrate OK
judgment hysteresis
(1 % FS)
*
11
(0.5 to 100 % FS)
*
11
3
A.
HI
*
2
,
*
3
Flowrate deviation
upper limit alarm
(10 % FS)
*
11
(0.5 to 100 % FS)
*
11
C-07
,
C-08
,
C-
15
,
C-
16
4
A.H.Hy
*
2
,
*
3
Flowrate deviation
upper limit alarm
hysteresis
(2 % FS)
*
11
(0.5 to 100 % FS)
*
11
5
A.
LO
*
2
,
*
3
Flowrate deviation
lower limit alarm
(10 % FS)
*
11
(0.5 to 100 % FS)
*
11
6
A.L.Hy
*
2
,
*
3
Flowrate deviation
lower limit alarm
hysteresis
(2 % FS)
*
11
(0.5 to 100 % FS)
*
11
7
A.dLy
*
3
Flowrate deviation
alarm judgment
delay time
10.0 s 1.0 to 999.9 s
8
E.
1.dL
*
4
Event output 1
delay time
0.0 s 0.0 to 999.9 s
C-07
,
C-08
Even if the delay time is set, it is dis-
abled during selection of integration
pulse output.
9
E.2.dL
*
4
Event output 2
delay time
0.0 s 0.0 to 999.9 s
10
C.F.
*
5
User setup conver-
sion factor
1.000 0.100 to 9.999
C-
18
11
E.
1.SP
*
6
Event output 1
upper / lower limit
owrate setup
(0 % FS)
*
11
(0 to 100 % FS)
*
11
C-07
,
C-08
Unit: L / min(standard)
12
E.2.SP
*
6
Event output 2
upper / lower limit
owrate setup
(0 % FS)
*
11
(0 to 100 % FS)
*
11
13
A.SCL
*
7
Analog optional
scaling
(100 % FS)
*
11
(10 to 100 % FS)
*
11
C-28
The owrate of analog input / output
100 % (5 V) is set.
Unit: L / min(standard)
14
E.A.LO
*
8
Integrated event
owrate
(lower 4 digits)
0 0 to 9999
C-07
,
C-08
,
C-
13
15
E.A.HI
*
8
Integrated event
owrate
(upper 4 digits)
0 0 to 9999
16
P.0.dL
*
9
PV forced zero
function
3.0 s 0.0 to 999.9 s
C-29
17
S.P.L.H
*
10
SP upper limit ow-
rate
(100 % FS)
*
11
(10 to 100 % FS)
*
11
C-35
Unit : L/min (standard)
When the upper limit owrate is set lower
than the lower limit owrate, the upper
limit owrate is prioritized than the lower.
18
S.P.L.L
*
10
SP lower limit ow-
rate
(0 % FS)
*
11
(10 to 100 % FS)
*
11
*1 Operation during judgment of flowrate OK.
SP
PV
ON(OK)
OFF
0.HyS
0.HyS
*2 Operation during judgment of flowrate deviation upper and lower limit alarms.
SP
PV
ON(ALM)
OFF
A.LoA.HI
A.L.Hy
A.H.Hy