Technical Data Hi Flow Sampler
Instruction 0055-9017
6
Natural Gas Sensor:
Detection Method...............Catalytic oxidation / Thermal conductivity
Range:
Catalytic oxidation.........0 to 5% by volume methane
Thermal conductivity .....5 to 100% by volume methane
Accuracy............................±5% of reading or 0.02 % methane, whichever is greater
Battery:
Type ..................................Intrinsically Safe NiMH rechargeable pack
Voltage ..............................4.8 V, max.
Recharge Time ...................8 to 10 hours
Run Time...........................>4.5 hours continuous operation @ 20 °C (68 °F)
Memory ................................Stores up to 1000 individual test parameters
Dimensions ..........................18L x 12W x 7H inches (457 x 305 x178 mm)
Weight ..................................20 lbs (0.9 kg)
Agency Approvals .................Designed to be intrinsically safe for use in hazardous locations
Class I, Division 1, Groups A, B, C & D in North America
CAN/CSA-C22.2 No. 157 – (June 1992)
ANSI (June 27, 2002)/UL913-2002
Intrinsically Safe Apparatus and Associated Apparatus for use in
Class I, Division 1, Hazardous (Classified) Locations
CE Mark – See this manual’s inside front cover for Declaration of
Conformity statement