Operating Manual BaehrTec A1200 / A2000
Page 48/64 English – V100 - Date 09/2017
Instruments
do not
• Maximum permitted speed
for instrument in use is
exceeded
• Instrument shaft worn
• Instrument shaft bent
• Instrument not correctly
mechanism
• Cream, ointment, or similar
on instrument shaft
• Too much work pressure on
instrument
•
Pulling movement while
working
• Only work with the
maximum permissible
speed for the instruments
used
• Check the instrument shaft
and use a new instrument
if necessary
•
without fail
• Insert instrument until stop
in clamping mechanism
•
with small instruments until
larger instruments start to
clamp securely again.
During operation, always
make sure that the
handpiece tip and the
instruments are clean.
• Reduce pressure
• Avoid pulling
Suction turbine
switches to
The dust bag, turbine protection
filter and filter cartridge are full, the
device will automatically switch to
suction level 3 to prevent damage
due to a rise in the temperature
Change all filters and the dust
bag and make sure that the
extracted air can escape easily
from the back of the device.
Handpiece
vibrates;
loud with used
instruments
• Maximum permitted speed
for instrument in use is
Instrument defective (bent,
shaft worn, etc.)
• Only work with the
maximum permissible
speed for the instruments
used
• Use new instruments.
Handpiece or handpiece button
soiled
Clean the push-button
If one of these faults still persists despite having followed this information, please send the
unit to the customer service team.
In addition, please get in touch with the manufacturer for each fault / each malfunction in cases
of doubt.