Operating Manual BaehrTec A1200 / A2000
English – V100 - Date 09/2017 Page 47/64
5.5 Self-help in the event of malfunctions
Fault
Cause
Solution
Control unit
not
working
•
Main switch off
• Control unit not connected to
power supply
•
Connecting outlet without
power
•
Mains cable not contacted
correctly (plugged in)
•
Fuse(s) problem (see
section 5.5.1 Replacing the
fuse)
•
Turn on main switch
• Connect device to power
supply
• Connect the device to a
functional socket
•
Plug device connector
correctly into the device
socket
•
Check main fuse and
replace if necessary
The device m
ust be
switched off and
disconnected from the
power supply before
checking the fuses.
Fault
Cause
Solution
Filter indicator
lit
• Handpiece clogged
• Dust bag is full
•
Turbine protection filter in
suction box is blocked
•
Filter cartridge is blocked
• Unscrew the front panel
and clean the handpiece.
It is essential that the
device is turned off and
disconnected from the
mains beforehand!
• Change dust bag
• Change turbine protection
filter
•
Change filter in cartridge
Low
suction power
and
the device gets
hot
•
Filter indicator ignored
• Handpiece clogged
• Dust bag is full
•
Turbine protection filter in
suction box is blocked
• Filter cartridge is blocked
• Exhaust is blocked
• Dust bag lid is not properly
closed
• The seal from the guide on
the control u
nit or on the
handpiece outlet has
slipped, is dirty or defective
•
Regularly check for the
filter check display
•
Unscrew the front panel
and clean the handpiece.
It is essential that the
device is turned off and
disconnected from the
mains beforehand!
• Change dust bag
• Change turbine protection
filter
• Change filter in cartridge
•
Please ensure that the
extracted air can escape
easily from the back of the
device
• Properly close the dust bag
lid
• Check O-rings and replace
as required. Clean filter
box, seal channel and dust
bag cover.