Operating Manual BaehrTec A1200 / A2000
English – V100 - Date 09/2017 Page 47/64
5.5 Self-help in the event of malfunctions
Control unit
not
working
•
• Control unit not connected to
power supply
•
Connecting outlet without
Mains cable not contacted
section 5.5.1 Replacing the
fuse)
•
• Connect device to power
supply
• Connect the device to a
functional socket
•
Plug device connector
correctly into the device
replace if necessary
The device m
ust be
switched off and
disconnected from the
power supply before
Turbine protection filter in
Filter cartridge is blocked
• Unscrew the front panel
mains beforehand!
• Change dust bag
• Change turbine protection
filter
•
Change filter in cartridge
Low
suction power
and
the device gets
hot
•
• Handpiece clogged
• Dust bag is full
•
Turbine protection filter in
suction box is blocked
• Filter cartridge is blocked
• Exhaust is blocked
• Dust bag lid is not properly
closed
• The seal from the guide on
the control u
nit or on the
handpiece outlet has
slipped, is dirty or defective
•
mains beforehand!
• Change dust bag
• Change turbine protection
filter
• Change filter in cartridge
•
extracted air can escape
easily from the back of the
device
• Properly close the dust bag
lid
• Check O-rings and replace
as required. Clean filter
box, seal channel and dust
bag cover.