Operating Manual BaehrTec A1200 / A2000
English – V100 - Date 09/2017 Page 7/64
Permitted temperature range: -10°C to +40°C
Permitted humidity range: 30% to 85%
Permitted air pressure: 800 hPa - 1060 hPa
2.1.4 Differences between BaehrTec A1200 and BaehrTec A2000
Speed range handpiece motor
The speeds (rpm) of the handpiece motor can be selected directly on the front controls
Setting the speeds (rpm) of the handpiece motor using the buttons (27) and (28) (Fig. 4a
and 4b)
1,000 rpm increments (6,000 to 31,000)
1,000 rpm increments from 6,000 to
40,000
-30 mbar (setting 1) to -70 mbar (setting 6)
-35 mbar (setting 1) to -100 mbar (setting
6)
2 memory buttons (Fig. 4a (29) + (30))
3 memory buttons (Fig. 4b (29) + (30) +
(31))
This device cannot be operated
by foot pedal. There is no connection on
the back of the device
This device can be operated by foot pedal.
There is a connection on the back of the
device
For details about the speeds of the handpiece motor and suction please see chapter
"Technical Data".
Please particularly note the differences between the BaehrTec A1200
and BaehrTec A2000 when using instruments and when adjusting
the handpiece speed.
All descriptions related to the foot pedal apply to the BaehrTec A2000 only,
not the BaehrTec A1200.