Operating Manual BaehrTec A1200 / A2000
English V100 - Date 09/2017 Page 7/64
Protect from wetness!
Permitted temperature range: -10°C to +40°C
Permitted humidity range: 30% to 85%
800 hPa - 1060 hPa
Permitted air pressure: 800 hPa - 1060 hPa
2.1.4 Differences between BaehrTec A1200 and BaehrTec A2000
BaehrTec A1200
Speed range handpiece motor
6,000 35,000 rpm
The speeds (rpm) of the handpiece motor can be selected directly on the front controls
6,000
10,000
15,000
20,000
25,000
30,000
32,500 (Fig. 4a (11))
35,000 (Fig. 4a (12))
Setting the speeds (rpm) of the handpiece motor using the buttons (27) and (28) (Fig. 4a
and 4b)
1,000 rpm increments (6,000 to 31,000)
1,000 rpm increments from 6,000 to
40,000
31,000

32,500
32,500

34,000
34,000

35,000
Maximum vacuum
-30 mbar (setting 1) to -70 mbar (setting 6)
Memory buttons
2 memory buttons (Fig. 4a (29) + (30))
Foot pedal operation
This device cannot be operated
by foot pedal. There is no connection on
the back of the device
(see Fig. 9 (48))
There is a connection on the back of the
device
For details about the speeds of the handpiece motor and suction please see chapter
"Technical Data".
Please particularly note the differences between the BaehrTec A1200
and BaehrTec A2000 when using instruments and when adjusting
the handpiece speed.
All descriptions related to the foot pedal apply to the BaehrTec A2000 only,
not the BaehrTec A1200.