Operator’s Manual - Rev. C - 05/05
- 26 –
6.3 TEMPERATURE CALIBRATION
1. [MODE] to go to
_
D2 |TEMP.: 20.0 °C |
2. [CAL] to access the calibration procedure
_
|CAL T 20.0°C |
'>>>>>' ('<<<<<'): Temperature value overrange
[MODE] to stop the procedure and to go back to D2
[/\]+[\/]+[ENT] to press the three keys to turn to factory calibration
3. [/\] [\/] to modify the actual reading
4. [ENT] to confirm and to go to the manual Temperature adjustment
MESSAGE FUNCTION
_
| " UPDATE " | The calibration is accepted
Error messages
_
| Z> 2.0°C | Zero > 2.0 °C
The above message will last for 5 minutes
[ENT] to acknowledge the message
_
|" NO UPDATE " | calibration not accepted
Manual Temperature calibration
_
|CAL T.M: 20.0°C|
[MODE] to stop the procedure and to go back to D2
1. [/\] [\/] to modify the actual value
2. [ENT] to confirm and to go back to D2