B&C Electronics CL 7685
Operator’s Manual - Rev. C - 05/05
- 35 –
7.12 ANALOG OUTPUT N°1 RANGE
_ _
|CAL OUT1: 0/20mA| |CAL OUT1: 4/20mA|
0/20mA (4/20mA): range selected
Active keys: [MODE] - [/\] [\/] - [ENT]
_
|CAL P1: 0.000 ppm|
P1: begin of range
0.000 PPM: measuring value related to 0/4 mA
Active keys: [MODE] - [/\] [\/] - [ENT]
_
|CAL P2: 2.000 ppm|
P2: end of range
2.000 PPM: measuring value related to 20 mA
Active keys: [MODE] - [/\] [\/] - [ENT]
IMPORTANT NOTE 1: if the value related to P1 is higher than the value related to P2 the analog
output will be the "reverse", otherwise will be the "direct" type.
The display will show OUT2 instead of OUT1 and the operator will follow the same procedure for the
output n°2 if the option 091.3711 dual output is installed.
7.13 INPUT RELATED TO ANALOG OUTPUT N°2 (OPTION 091.3711)
_
|CAL OUT2: ppm
| (°C)
PPM: input selected for analog output N°2
Active keys: [MODE] - [/\] [\/] - [ENT]