9. Operating Instructions – Change Program
9-44 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
9.8.4 Hi-Dose CCPD Calculated Settings
HomeChoice/HomeChoice PRO APD System calculates the number of night cycles
and the Dwell Time. The calculated values are shown on the display screen when
you press
STOP after you have completed reviewing your therapy settings.
Table 9-14. Definitions for Hi-Dose CCPD Calculated Settings
Setting Description
Cycles Total number of cycles at night, not including the Last Fill.
This parameter is calculated by the system.
Dwell Time Calculated amount of time the dialysis solution remains in the
peritoneal cavity during each cycle. The system may
automatically adjust the Dwell Time based on your actual
flow rates during Fill and Drain, if your clinician chooses this
option.