07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 19-1
Technical Data
The specifications in this section apply to the Baxter HomeChoice and HomeChoice
PRO APD Systems.
19.1 Physical Specifications
19.2 Electrical Power Requirements
Line voltage disturbances or improper grounding can adversely affect operation of the
HomeChoice and HomeChoice PRO APD Systems. Variations in line voltage amplitude
should be less than ±10 percent of nominal voltage. The duration of any voltage
disturbance should be less than 5 milliseconds.
U.S. Standard Metric
Height: 7.0 in 17.8 cm
Width: 19.5 in 49.5 cm
Depth: 15.7 in 39.9 cm
Weight: 27 lbs 12.3 kg
Product codes: 5C4471, 5C4471R, 5C8310, 5C8310R
Voltage range: 115 VAC ±10%
Frequency range: 50/60 Hz
Fuses: F 5.0A – 125V
Mode of operation: Continuous
Degree of protection against electrical shock: Type B Applied Part
Type of protection against electrical shock: Class I equipment
Power consumption: Maximum: 600 VA (600 watts)
Average: 100 VA (100 watts)
Degree of protection against ingress of water: Meets the requirements of
IEC 60601-2-39 (Clause 44.3)