EasyManua.ls Logo

baxter HomeChoice - Attach the Drain Option

baxter HomeChoice
377 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11. Operating Instructions – Prepare for Therapy
11-14 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
11.6 Attach the Drain Option
Follow the steps below to attach your drain option.
WARNING
When a drain line extension is used instead of a drain bag,
you must leave an air gap (space) between the end of the
drain line and any fluid in the drain or container. This
prevents non-sterile fluid from flowing backwards up the
drain line. Non-sterile fluid can contaminate the fluid path
resulting in peritonitis.
Steps to attach a drain option
1. Attach your drain option:
Drain Line Extension – remove tip
protectors from both ends of drain.
– OR –
15L Drain Bag – close clamp on short tube
to prevent leakage.
h NOTE: If you use more than one drain bag, use
a drain manifold to connect the bags.
2. Open all clamps in the drain lines.
3. Press
GO.
The display screen changes to SELF TESTING.
When the self-test is complete, CONNECT BAGS
appears.
SELF TESTING...
CONNECT BAGS

Table of Contents