18. Troubleshooting
18-20 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.3.9 Machine Tilted
MACHINE TILTED
Display Message:
Cause: The cycler is tilted.
If this alarm occurs during therapy, it is a Auto Restart
Alarm.
To Correct: 1. Make sure that the system is on a flat surface and is not
tilted.
– OR –
If the alarm becomes a Manual Restart Alarm or occurs at
PRESS GO TO START:
1. Press
STOP to mute the alarm.
2. Place the cycler on a flat, even surface.
3. Press
GO to return to the therapy.
MACHINE TILTED