07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 13-11
13. Operating Instructions – End Therapy
Table 13-1. End of Therapy Options
Option Description
Initial Drain
Volume
Volume from Initial Drain of the current therapy.
Last M-Drain
Only appears if you drained some fluid using the Manual
Drain option before a Last Fill. Shows the amount of
solution drained.
Total
Ultrafiltration
The total amount of UF for the entire night’s therapy. Press
ENTER to review cycle-by-cycle information. Press STOP
to exit cycle-by-cycle information.
h NOTE: A lower than usual or negative Total UF at the
end of a therapy may indicate that your last
Drain was incomplete. The Last Manual Drain
option should be turned ON and a UF Target
programmed with a value that equals around
70% of your expected UF. For help in
converting 70% of your expected total therapy
UF into a value that can be programmed as
your Total UF for your Tidal therapy, see
19.17, Determining Tidal Total UF and Last
Manual Drain UF Target Volume Settings, on
page 19-18. See also 10.2.9, Last Manual Drain,
on page 10-14.
Average Dwell
Time
Average actual Dwell Time for your therapy. Press
ENTER
to review cycle-by-cycle information. Press STOP to exit
cycle-by-cycle information.
I-DRAIN VOL: 65ML
LAST M-DRAIN: 60ML
TOTAL UF: 150ML
AVG DWELL TIME: 1:32