18. Troubleshooting
18-44 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
Steps to bypass
Drain Not Finished alarm Display screen
The DRAIN NOT FINISHED alarm appears when you try
to bypass and your current Drain Volume is less than
the Minimum Drain Volume.
1. Press
STOP to silence the alarm.
2. Press
V.
The volume drained is subtracted from your Fill
Volume. Your next Fill is reduced by this difference.
If this alarm occurs during Initial Drain,
DRAIN VOLUME: ML does not appear.
3. Press
V.
The Initial Drain Volume (I-DRAIN VOL) from the
current therapy is shown.
4. Press
V.
The TOTAL UF from the end of the previous Drain
is shown.
5. Press
V.
6. Press
ENTER to bypass the DRAIN NOT FINISHED
alarm.
The next FILL begins. The volume of fluid that was
not drained is subtracted from the Fill Volume for
this phase.
DRAIN NOT FINISHED
DRAIN VOLUME: ML
I-DRAIN VOL: ML
TOTAL UF: ML
BYPASS
FILL 3 OF 5