18. Troubleshooting
18-48 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
h NOTE: If LOW DRAIN VOLUME alarms continue to occur after you resume
Drain, and it is necessary to bypass the Drain (for example, to reverse
the flow in the catheter to correct a fibrin blockage issue):
Press STOP to silence a LOW DRAIN VOLUME alarm.
Press GO to resume Drain.
Press STOP to stop the Drain (when no alarm is present).
Bypass the Drain phase. See 18.4.3, Bypass Drain Phase, on
page 18-41.
A DRAIN NOT FINISHED alarm occurs if you have not met your
minimum Drain Volume when bypassing a Drain phase. If a DRAIN
NOT FINISHED alarm occurs, your next Fill will be reduced by the
amount that your Drain Volume was less than your prescribed Fill
Volume. If you do not encounter a DRAIN NOT FINISHED alarm, your
next Fill Volume will equal your prescribed Fill Volume.
Steps to bypass
Low Drain Volume alarm (Continued) Display screen