User’s Guide44
Base configurations Each flow cytometer has a base cytometer configuration that
corresponds to the layout of the installed lasers and optics in your
cytometer. This base configuration is set by your field service
engineer.
BD FACSDiva
cytometer
configuration
Before you acquire data using BD FACSDiva software, you must
specify a cytometer configuration. The cytometer configuration
defines which filters and mirrors are installed at each detector.
BD FACSDiva software provides a BD base configuration for your
flow cytometer. Select Cytometer > View
Configuration to create,
modify, or delete custom cytometer configurations. (See the
Cytometer and Acquisition Controls chapter of the BD
FACSDiva
Software Reference Manual for details.)