User’s Guide78
Creating application settings
Introduction This topic describes how to create application settings.
About application
settings
Application settings are associated with a cytometer configuration
and include the parameters for the application, area scaling values,
PMT voltages, and threshold values, but not compensation. Each
time a performance check is run for a configuration, the
application settings associated with that configuration are updated
to the latest run.
Using application settings provides a consistent and reproducible
way to reuse cytometer settings for commonly used applications.
You can include area scaling adjustment in your application
settings. See
Adjusting area scaling (page 115) for more
information.
Before you begin Perform the cytometer setup procedure and run a performance
check for the configuration that will be used for the application.
Procedure To create application settings:
1. In the open experiment, right-click Cytometer Settings in the
Browser, then select Application Settings > Create
Worksheet.
A second global worksheet is added with the plots created
according to the selections in the Parameters tab.