BD Veritor™ Plus Analyzer Instructions for Use
34
13
ANALYZER
FAILURE
CALL BD
ERROR 13
A hardware anomaly has
occurred.
Contact BD*.
14
ANALYZER
FAILURE
CALL BD
ERROR 14
A hardware anomaly has
occurred.
Contact BD*.
15
ANALYZER
FAILURE
CALL BD
ERROR 15
A hardware anomaly has
occurred.
Contact BD*.
16
ANALYZER
FAILURE
CALL BD
ERROR 16
A hardware anomaly has
occurred.
Contact BD*.
17
ANALYZER
FAILURE
CALL BD
ERROR 17
A hardware anomaly has
occurred.
Contact BD*.
18
ANALYZER
FAILURE
CALL BD
ERROR 18
A hardware anomaly has
occurred.
Contact BD*.
19
BACKGROUND INVALID
CONSULT MANUAL
ERROR 19
The assay background area
is outside of the normal
range.
Remove the device and cycle the instrument power.
If the instrument displays an error during power-on,
contact BD*.
If no error message appears, obtain an additional
sample and repeat the assay If the issue recurs,
contact BD*.
20
ANALYZER
FAILURE
CALL BD
ERROR 20
A hardware anomaly has
occurred.
Contact BD*.
Result Number Message Possible Cause(s) Corrective Action(s)
* outside of the United States contact your local BD representative