35
21 / 22
ANALYZER
EXPIRED
NO DAYS
REMAIN
The shelf life of the
BD Veritor™ Plus Analyzer
has expired. The maximum
shelf-life of the BD Veritor™
Plus Analyzer from the date
of manufacture is 34 months
(regardless of time of first
use).
or
The operational life of the
BD Veritor™ Plus Analyzer
has expired. The maximum
number of months from first
use is 24 months.
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
23
ANALYZER
EXPIRED
NO TESTS
REMAIN
The operational life of the
BD Veritor™ Plus Analyzer
has expired. The maximum
number of tests performed by
a BD Veritor™ Plus Analyzer
is 10,000 runs.
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
Outside of the United States contact your local BD
representative
24
INVALID BARCODE
ON TEST DEVICE
REPEAT TEST WITH
NEW TEST DEVICE
The test device barcode
cannot be read correctly.
Repeat the assay. On subsequent occurrences, contact
BD*.
25
VERIFY FAIL
CONSULT
MANUAL
ERROR 25
The result of a BD Veritor™
System Verification Cartridge
run is unsuccessful.
Ensure that no foreign matter is present on the
BD Veritor™ System Verification Cartridge assay area,
and repeat the run. On subsequent occurrences, contact
BD*.
26
RESULT INVALID
FLU A: + FLU B: +
CONFIRM RESULT
ERROR 26
The BD Veritor™ Plus
Analyzer has detected
potential interference in the
test line area.
Specimens generating a “Result Invalid” with Error 26
should be retested. Upon retesting, if the specimen
produces a “Result Invalid” the user may want to consider
other methods to determine whether the sample is
positive or negative for influenza virus.
27
ANALYZER
FAILURE
CALL BD
ERROR 27
A hardware anomaly has
occurred.
Contact BD*.
28
ANALYZER
FAILURE
CALL BD
ERROR 28
A hardware anomaly has
occurred.
Contact BD*.
29
ANALYZER
FAILURE
CALL BD
ERROR 29
A hardware anomaly has
occurred.
Contact BD*.
Result Number Message Possible Cause(s) Corrective Action(s)
* outside of the United States contact your local BD representative