37
none
INVALID
KIT LOT
BARCODE
An invalid kit lot barcode was
detected by the BD Veritor™
Plus Analyzer.
Ensure that the barcode scanned during this step is from
a valid BD Veritor™ Test Kit box label. For kits with two
barcodes, scan only the upper barcode that starts with
(17).
none
OPERATOR ID
SCAN TIMEOUT
REINSERT
TEST DEVICE
No barcode was scanned
within 30 seconds.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
none
SPECIMEN ID
SCAN TIMEOUT
REINSERT
TEST DEVICE
No barcode was scanned
within 30 seconds.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
none
LOT NUMBER
SCAN TIMEOUT
REINSERT
TEST DEVICE
No barcode was scanned
within 30 seconds.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
none
XXX DAYS
ANALYZER LIFE
REMAINING
The shelf or operational life
of the BD Veritor™ Plus
Analyzer is nearing
expiration. The maximum
shelf-life of the BD Veritor™
Plus Analyzer from the date
of manufacture is 34 months
(regardless of time of first
use). The maximum number
of months from first use is
24 months.
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
none
XXX READS
REMAINING
The BD Veritor™ Plus
Analyzer is nearing its usage
expiration. The maximum
number of tests performed by
a BD Veritor™ Plus Analyzer
is 10,000 runs.
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
none
Flashing Plug icon
Failure of the internal battery.
Battery Not Charging:
In the event the battery icon
does not indicate the battery
is charging while the power
supply is plugged into the
BD Veritor™ Plus Analyzer,
examine the AC Adapter
plugged into the Analyzer.
The supplied AC Adapter
has a BD label near the
coaxial power connector and
conforms to the sample
labeling in section 2.3.
Contact BD*.
Note: the Analyzer may continue to be used while
plugged into a facility power source.
Result Number Message Possible Cause(s) Corrective Action(s)
* outside of the United States contact your local BD representative