LB 124 Scint 6. Software Functions
69
Note:
If the device is placed into the wall holder or if it is turned on,
the rechargeable batteries inside the device can be charged au-
tomatically, provided the Charge mode is set to ON. Via the
menu item Charge time the duration of the charging process is
limited to max. 12 hours to prevent overcharging. If you take out
the device from the wall holder and place it into the wall holder
again later, or if you turn the device off and on again, then the
charging process will be re-started, provided the Charge mode
is enabled.
Caution:
The charge function is active only if the device has
been turned on.
Æ Charge Mode Here you can set the charge mode (ON/OFF). Only with setting
Accu.
Æ Charge time Here you set the duration of the charging process. Only with
setting Accu.
Æ Line Voltage Here the current line voltage is indicated. This value is deter-
mined and displayed by the device.
Æ Voltage Battery Here the current battery voltage (or accu voltage) is indicated.
Please take into account that the battery voltage indication is
not relevant as long as the mains voltage is being supplied, as
the battery is not in use during this time.
Æ System Timeout Here an automatic switch-off function for the LB 124 Scint can
be activated. When this function is activated, the device is au-
tomatically switched off when it is not operated for a predefined
period of time, i.e. when no key is pressed. The default setting
is 0 (function disabled).
This function is not enabled during a charging process!
6.8.11 Hardware
Figure 50: Hardware settings
These displays are not relevant for the user.