6. Software Functions LB 124 Scint
70
6.9 Profile
6.9.1 Overview
The access rights for different user (groups) can be defined on
this menu. For each user, only the menu items will appear
which he is allowed to work with (read and/or edit). At most 4
profiles are possible. Three are already predefined and can be
modified.
EASY Only measurements are possible: gross, cps, ratemeter
STANDARD Here you can choose the menu items which are important for
standard situations: All nuclides can be chosen, the small nu-
clide library is preset with gross, net and Co-60, the calibration
factors however cannot be edited. As measurement mode, you
can select ratemeter, scaler-timer and survey.
EXPERT All possibilities are given, the parameters can be edited (incl.
the calibration factors).
Status after initial start-up
After the initial start-up, the device switches to the ratemeter
mode and the STANDARD profile.
In this profile, only simple measuring tasks are accessible
ratemeter or background measurement modes, nuclide selec-
tion from Net or ß-Tot, as well as switching between the lan-
guages. Under the profile menu item you can switch to the pro-
file EXPERT. In this profile, all functions are accessible (see
section 6.6.1).
The name of the currently selected profile is indicated in the top
line of the display.
When the device is switched on, it loads the measurement
mode predefined in the Profile menu (the factory-setting is the
standard profile). The profile that was active when the device
was last switched off will be automatically loaded.
The user can select a different profile, if he knows the appropri-
ate PIN number required.