Alarms
Precellys
®
24
Page 25/35
7.2.2 Cover Alarm
If the cover is opened while a run is in progress, the cover alarm sets off and the run stops
immediately. The following message is displayed on the screen:
TOP ALARM
PUSH VALID
Red light is flashing / green light is off
To return to the main menu, the user must press the “Valid” button.
Actions to be done in case of alarm are described in § 7.3.
7.2.3 Speed Error Alarm
If the actual speed is different from the assigned speed, the speed error alarm sets off and the run
stops immediately. The following message is displayed on the screen:
ERR SPEED SENSOR
PUSH VALID
Red light is flashing / green light is off
To return to the main menu, the user must press the “Valid” button.
Actions to be done in case of alarm are described in § 7.3.