BF (E1) 05/2016 page 42/58
Electrical data
(model versions BF053-230V, BF115-230V, BF240-230V, BF400-230V, BF720-230V)
IP system of protection acc. to EN 60529
voltage
at 50 Hz power frequency
V 230 230 230 230 230
at 60 Hz power frequency
V 230 230 230 230 230
Chamber fuse 5 x 20 mm
230V / 10A / middle-time-lag (M)
Installation category acc. to IEC 61010-1
Pollution degree acc. to IEC 61010-1
Different electrical data for BF-UL constructed for the USA and Canada
(model versions BF053UL-120V, BF115UL-120V, BF240UL-120V, BF400UL-120V, BF720UL-120V)
Nominal voltage (±10 %)
at 60 Hz power frequency
V 115 115 115 115 115
Chamber fuse 6,3 x 32 mm
250V / super-time-lag TT
Additional temperature protection
class 1 (DIN 12880:2007)
internal internal internal internal internal
Environment-specific data
Energy consumption at 37 °C / 98.6 °F
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C ± 3 °C / 71.6 °F ± 5.4 °F and a power supply voltage fluctuation of ±10. Technical data is
determined in accordance to BINDER Factory Standard Part 1:2015 following DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
12.5 Equipment and options (extract)
To operate chamber, use only original BINDER accessories or accessories / components from
third-party suppliers authorized by BINDER. The user is responsible for any risk arising from
using unauthorized accessories.
Microprocessor temperature controller with LED display and several time functions
Three Timer functions: Delayed On, Delayed Off and temperature dependent delayed Off
Temperature safety device class 3.1 acc. to DIN 12880:2007 with visual temperature alarm
Rear exhaust duct, internal diameter 50 mm / 1.97 in with ventilation slide
Adjustable air change by rear exhaust duct (50 mm / 1.97 in ) with ventilation flap and front ventilation
slide
Four castors, 2 lockable (size 720 only)
RS 422 interface for communication software APT-COM™ DataControlSystem, or switch over to printer
output with RS 232/RS 422 interface converter