20.4 KB 23 / KB 23-UL (E3.1) technical data
Exterior dimensions
Width, net
mm / inch
435 / 17.13
Height, gross (incl. feet)
mm / inch
620 / 24.41
Depth, net
mm / inch
520 / 20.47
Depth, gross (incl. door handle, I-panel, connection, and 30
mm for cable)
mm / inch
600 / 23.62
Wall clearance (minimum)
mm / inch
100 / 3.94
Wall clearance door open (minimum)
mm / inch
100 / 3.94
Doors
Number of doors
1
Number of inner glass doors
1
Interior dimensions
Width
mm / inch
222 / 8.74
Height
mm / inch
330 / 12.99
Depth
mm / inch
277 / 10.91
Interior volume
l / cu.ft.
10 / 0.35
Steam space volume
l / cu.ft.
Racks
Quantity of racks (regular)
2
Quantity of racks (max.)
3
Maximum load per rack
Kg / lbs
12 / 26
Maximum permitted total load
Kg / lbs
25 / 55
Weight
Weight (empty)
Kg / lbs
44 / 97
Temperature data
Temperature range
°C
°F
0 to +100
4 to 212
Temperature fluctuation
at 5 °C
± K
0.2
at 40 °C
± K
0.2
Temperature uniformity (variation)
at 5 °C
± K
0.7
at 25 °C
± K
0.3
at 40 °C
± K
0.3
Heating up time
to 40 °C
min
11
Cooling down time from 40 °C
to 5 °C
min
61
Recovery time after doors
were open for 30 sec.
at 5 °C
min
5
at 40 °C
min
2
Electrical data (model version KB023-230V)
IP system of protection acc. to EN 60529
20
Nominal voltage (±10 %) at 50 Hz power frequency
V
230
Current type
1N~
Power plug
Grounded plug
Nominal power
kW
0.34
Overvoltage category acc. to IEC 61010-1
II
Pollution degree acc. to IEC 61010-1
2
Chamber fuse 5x20 mm 230V, time-lag T, external
Amp
10