KMF (E5.2) 06/2014 page 85/99
Fault description
Possible cause
Required measures
Controller (continued)
Program duration longer than
programmed.
Tolerances have been pro-
grammed.
For rapid transition phases, do
NOT program tolerance limits in
order to permit maximum heating,
refrigerating, or humidification
speed.
Program stops one section too
early.
Program line is incomplete.
When programming, define the
end value of the desired cycle by
adding an additional section with
a section time of at least one
second.
Ramp temperature transitions are
only realized as steps.
When using the Program Editor
of the software APT-COM™ 3
DataControlSystem, the setting
“step” has been selected.
Select setting “ramp” in the Pro-
gram Editor of the software APT-
COM™ 3 DataControlSystem
and transfer a program to the
chamber controller.
Humidity alarm message when
operating without humidity (hu-
midity switch (4) OFF)
Humidity set-point set to a value
> 0% r.F.
Manual Mode: Enter a humidity
set-point 0% r.H.
Program Mode: Enter a humidity
subprogram with humidity set-
point 0% r.H.
RESETbutton does not cancel
the notifying or alarm indication.
Cause of disturbance not re-
moved correctly
The RESETbutton permits
resetting notifying or alarm mes-
sages for temperature and hu-
midity only with in a tolerance
sector of +/- 2 °C resp. +/- 5 %
r.H.
Remove cause of disturbance. If
the RESETbutton still does not
cancel the indication, contact
BINDER service.
Display flashing:
1999 or -1999 or 9999.
Sensor rupture between sensor
and controller or Pt 100 sensor
defective.
Contact BINDER service.
Short-circuit.
Initialization problem due to turn-
ing on the chamber too early.
Observe a delay time of about
30s between turning the chamber
Off and On again.
Miscellaneous
Impaired valve function of hose
burst protection.
Calcification.
Remove calcifications by citric
acid or acetic acid solutions
(chap.
4.3.4).
Have a plumber inspect the
valve.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired units must comply with the BINDER quality standards.