LifePro Thermal Cycler Instruction for operation
11
Press character key
in this interface to input
file name, press confirmation key
to move
the cursor forward. Select User and File
Name by up/down key
. Delete character by
left key
, move cursor by right key. Press
Caps Key
to shift between capital and lower
case. At upper left on screen it indicates
capital or lower case (as shown in the figure
abc).
Press Main” key to return to main
interface
Press Save” key to save the file. When
file is saved normally, the system shows
“File have been saved. Please continue
another operation”. Press confirmation
key to make it disappear. If file has the
same name with another, the system
asks to name again (Rename) or continue to save (Save).
Press number key to set and save password.
Press Run” to access the interface for file running if Control Mode is Block now, press “Run” to
bounce the dialog box below if Control Mode is Tube now:
It is the same with main interface; confirm the password; input the sample type and micro liter
amount of sample size.
Caution: 1. File name and user name are not case sensitive.
2. Please remember the password.
3. The password should be 0~9 numeric characters.
As the file runs normally, “•••” at Now
Running keeps blinking. The system
shows “File run over” when running is
completed.
Press Stop”, it shows “Confirm stop
running?”
Select stop (stop) or continue (Start)
as per information.
Press Pause”, it shows “Now pause
running”, you can select continue
(Start) as per information.
Press View File” to view the edited
PCR program”.
Press Skip” to skip the current
temperature section and go to the
next temperature.
Hotlid=off 2009-05-1 10:05:00
abc
User PCR1
File Name PCR1
Password PC****
Password Check PC****
Main Caps Save Run
Interface for file saving
Select Sample Volume:
050
μl
File running interface 1
Hot lid=off 2009-05-1 10:05:00
User/File: Bioer/Bioer
Temp:
95.0C
Time:
01.00
Remain Time:
00h56m
Stop Pause Start Skip Run