Patient Monitor User’s Manual
7
Pulse width:0.1~2.0ms
Rise time:10~100us
Electrosurgery
protection
( electrotome
protection )
Cut power :300W
Coagulation power :100W
Change of heart rate : ≤10%
Recover time : ≤10s
Comply with the requirement of clause
4.2.9.14 in ANSI/AAMI EC 13:2002
Tall T-Wave
rejection
capability
Minimum recommended 1.2 mV T-Wave
amplitude. Comply with requirement of clause
4.1.2 c in ANSI/AAMI EC 13:2002.)
Heart rate
averaging
≤ 50 bpm, once every two beats;
50 bpm to 120 bpm, once every four beats;
> 120 bpm, once every six beats.
Heart rate
meter
accuracy and
response to
irregular
rhythm
Ventricular bigeminy : 80bpm
Slow alternating ventricular bigeminy : 60bpm
Rapid alternating ventricular bigeminy :
120bpm
Bidirectional systoles : 90bpm
Response time HR changes from 80bpm to 120bpm: less than